SECTION 9. INPUT/OUTPUT INSTRUCTIONS
9-6
The temperature value used in
compensating the RH value (Parameter 7)
must be obtained (see Instruction 11) prior
to executing Instruction 12.
The RH results are placed sequentially into the
input locations beginning with the first RH value.
In some situations the RH sensors might be
deployed such that only small temperature
variations exist within a given set of probes. In
these cases a single temperature may be used
to compensate the subset of RH measurements
instead of making a temperature measurement
for each RH probe. If the complete set of
temperature values are not needed, this
approach uses less input channels. Parameter
6 is used to specify how many consecutive RH
values get compensated per temperature
measurement.
In the 207 probe, the RH and temperature
elements use a common excitation line. Since
a single excitation channel can drive several
hundred probes, there is no
"measurements/excitation" parameter in
Instruction 12. NEVER EXCITE THE 207
PROBE WITH DC EXCITATION as the RH chip
will be damaged.
The maximum RH polynomial error is given
below:
Curve Fit Error --
Range (%RH)
Error (%RH)
10 - 100
±4
15 - 94
±1
PAR.
DATA
NO.
TYPE
DESCRIPTION
01:
2
Repetitions
02:
2
Analog card number for first
measurement
03:
2
Single-ended channel for first
measurement
04:
2
Excitation card number for
first measurement
05:
2
Excitation channel number
for first measurement
06:
2
Number of R.H.
measurements per
compensating temperature
measurement
07:
4
Input location for first
compensating temperature
measurement
08:
4
Input location for first
measurement
09:
FP
Multiplier
10:
FP
Offset
Input locations altered: 1 per repetition
*** 13 THERMOCOUPLE ***
TEMPERATURE, SINGLE ENDED
FUNCTION
This Instruction uses the selected thermocouple
calibration to calculate the thermocouple output
voltage at the reference temperature, then it
makes a SINGLE ENDED VOLTAGE
MEASUREMENT (Section 13.2) on the
thermocouple and adds the measured voltage
to the calculated reference voltage, then
converts the voltage to temperature in oC
(Section 13.4).
Table 9-3 gives the thermocouple type codes.
The reference temperature location will be
incremented by one each repetition if
C
is keyed
before entering Parameter 6. When this option
is exercised, two minus signs (--) will appear as
the right most characters of the display.
A multiplier of 1 and an offset of 0 yields
temperature in degrees C.
TABLE 9-3. Thermocouple Type Codes
Code
Thermocouple Type
1
T (copper - constantan)
2
E (chromel - constantan)
3
K (chromel - alumel)
4
J (iron - constantan)
5
B (platinum - rhodium)
6
R (platinum - rhodium)
7
S (platinum - rhodium)
1X
Output temperature difference between
Reference and Thermocouple
2X
Skip every other single ended channel
8X
TC input from A5B40 Isolation Amplifier
(use 5 V range)
Содержание CR7
Страница 2: ...This is a blank page ...
Страница 4: ...This is a blank page ...
Страница 28: ...CR7 MEASUREMENT AND CONTROL SYSTEM OVERVIEW OV 18 This is a blank page ...
Страница 38: ...SECTION 1 FUNCTIONAL MODES 1 10 This is a blank page ...
Страница 42: ...SECTION 2 INTERNAL DATA STORAGE 2 4 This is a blank page ...
Страница 52: ...SECTION 3 INSTRUCTION SET BASICS 3 10 This is a blank page ...
Страница 57: ...SECTION 4 EXTERNAL STORAGE PERIPHERALS 4 5 FIGURE 4 4 1 Example of CR7 Printable ASCII Output Format ...
Страница 58: ...SECTION 4 EXTERNAL STORAGE PERIPHERALS 4 6 This is a blank page ...
Страница 68: ...SECTION 6 9 PIN SERIAL INPUT OUTPUT 6 6 This is a blank page ...
Страница 140: ...SECTION 13 CR7 MEASUREMENTS 13 16 FIGURE 13 5 1 Circuits Used with Instructions 4 9 ...
Страница 162: ...APPENDIX A GLOSSARY A 4 This is a blank page ...
Страница 164: ...B 2 This is a blank page ...
Страница 170: ...APPENDIX C BINARY TELECOMMUNICATIONS C 6 This is a blank page ...
Страница 173: ...APPENDIX D CALIBRATION PROCEDURES D 3 FIGURE D 2 1 Calibration Points for the Analog Interface Card ...
Страница 174: ...APPENDIX D CALIBRATION PROCEDURES D 4 FIGURE D 2 2 CR7X CPU Card ...
Страница 186: ...CR7 INDEX I 8 This is a blank page ...