3-1
Section 3. CR5000 Measurement Details
3.1 Analog Voltage Measurement Sequence
The CR5000 measures analog voltages with either an integrate and hold or a
sample and hold analog to digital (A/D) conversion. The A/D conversion is
made with a 16 bit successive approximation technique which resolves the
signal voltage to approximately one part in 60,000 of the full scale range. The
maximum conversion rate is 5000 per second or one measurement every 200
µs (10,000 measurements per second on a single channel).
The timing of CR5000 measurements is precisely controlled. The measurement
schedule is determined at compile time and loaded into memory. This schedule
sets interrupts that drive the measurement task.
Using two different voltage measurement instructions with the same
voltage range takes the same measurement time as using one instruction
with two repetitions. (This is not the case in the CR10X, 21X, CR23X and
CR7 dataloggers where there is always a setup time for each instruction.)
There are four parameters in the measurement instructions that may vary the
sequence and timing of the measurement. These are options to measure and
correct the ground offset on single-ended measurements each time
measurements are made (MeasOfs), reverse the high and low differential
inputs (RevDiff), to set the time to allow the signal to settle between switching
to a channel and making a measurement (SettlingTime), and the length of time
to integrate a measurement (Integ), and to reverse the polarity of excitation
voltage (RevEx).
3.1.1 Voltage Range
The CR5000 has 5 fixed voltage ranges and autorange. The 16 bit A/D has a
resolution of 1 part in 2
16
(65,536). To allow for some overrange capabilities
the A/D is applied to a range approximately 9% greater than the Full Scale
Range resulting in the 1 part in 60,000 resolution over the FSR. For example,
on the
±
20 mV range the full scale range is 40 mV [20 - (-20)] and the
resolution is two thirds of a microvolt; 0.04 / 0.000000667 = 60,000. The
smaller the voltage range, the better the absolute resolution. In general, a
measurement should use the smallest fixed voltage range that will
accommodate the full scale output of the sensor being measured. If the voltage
exceeds the range, the CR5000 indicates the overrange by returning Not-A-
Number (NAN) for the measurement.
For signals that do not fluctuate too rapidly, AutoRange allows the CR5000 to
automatically choose the voltage range to use. AutoRange causes the CR5000
to make two measurements. The first measurement determines the range to
use. It is made with no integration on the
±
5000 mV range. The second
measurement is made on the appropriate range using the integration specified
in the instruction. Both measurements use the settling time programmed in the
instruction. AutoRange optimizes resolution but takes longer than a
measurement on a fixed range, because of the two measurements required.
Содержание CR5000
Страница 39: ...Section 1 Installation and Maintenance 1 15 S A N Y O FIGURE 1 11 6 Removal of band clamp and battery ...
Страница 40: ...Section 1 Installation and Maintenance 1 16 This is a blank page ...
Страница 54: ...Section 2 Data Storage and Retrieval 2 14 This is a blank page ...
Страница 88: ...Section 4 CRBasic Native Language Programming 4 12 This is a blank page ...
Страница 94: ...Section 5 Program Declarations 5 6 This is a blank page ...
Страница 172: ...Section 7 Measurement Instructions 7 46 This is a blank page ...
Страница 206: ......
Страница 236: ...Index Index 4 This is a blank page ...
Страница 237: ...This is a blank page ...