Section 7. Measurement Instructions
7-5
Parameter
& Data Type
Enter
RevDiff
Code Value
Result
(Reversing requires twice as much time to complete)
Constant
False 0
Signal is measured with the high side referenced to the low
True
≠
0
A second measurement is made after reversing the inputs to cancel
offsets
MeasOfs
Constant
Code Value
Result
the Ground offset voltage is subtracted from single
ended measurements.
False 0
Offset voltage is corrected from background calibration
True
≠
0
Offset voltage is measured each scan
SettlingTime
Constant
The time in microseconds to delay between setting up a measurement (switching to the
channel, setting the excitation) and making the measurement. (1 microsecond resolution)
Entry
Voltage Range
Integration
Settling Time
0
± 20 mV
Flash
200 µS (default)
0
All but ± 20 mV,
Flash
100 µS (default)
0
All
250 µS
200 µS (default)
0
All
_50Hz, _60 Hz
3 mS (default)
>=100
All All
µS
entered
Integ
The time spent on integration in microseconds for each of the channels measured.
Constant
Entry
Integration
0, 200
Flash Conversion; 200 = 2 flash conversions 100 µS apart
and averaged
250; 500-16000
250 µS; multiples of 500: 250 µS integrations starting each
500 µS and averaged
_60Hz or 16667
16,667 µS (reject 60 Hz noise)
_50 Hz or 20000
20,000 µS (reject 50 Hz noise)
Mult, Offset
Constant,
Variable,
Array, or
Expression
A multiplier and offset by which to scale the raw results of the measurement. See the
measurement description for the units of the raw result; a multiplier of one and an offset
of 0 are necessary to output in the raw units. For example, the
TCDiff
instruction
measures a thermocouple and outputs temperature in degrees C. A multiplier of 1.8 and
an offset of 32 will convert the temperature to degrees F.
7.3 Half Bridges
BrHalf (Dest, Reps, Range, SEChan, ExChan, MeasPEx, ExmV, RevEx,
SettlingTime, Integ, Mult, Offset)
X = result w/mult = 1,
offset = 0
X
V
V
R
R
R
x
s
s
f
=
=
+
1
This Instruction applies an excitation voltage, delays a specified time and then makes a
single ended voltage measurement. The result with a multiplier of 1 and an offset of 0 is
the ratio of the measured voltage divided by the excitation voltage.
Содержание CR5000
Страница 39: ...Section 1 Installation and Maintenance 1 15 S A N Y O FIGURE 1 11 6 Removal of band clamp and battery ...
Страница 40: ...Section 1 Installation and Maintenance 1 16 This is a blank page ...
Страница 54: ...Section 2 Data Storage and Retrieval 2 14 This is a blank page ...
Страница 88: ...Section 4 CRBasic Native Language Programming 4 12 This is a blank page ...
Страница 94: ...Section 5 Program Declarations 5 6 This is a blank page ...
Страница 172: ...Section 7 Measurement Instructions 7 46 This is a blank page ...
Страница 206: ......
Страница 236: ...Index Index 4 This is a blank page ...
Страница 237: ...This is a blank page ...