4. Press enter to make the measurement.
Table 8-4: RS-232 command list
Command
Values returned
Units
M0
1. Pressure
2. Temperature
3. Quality Metric
1. hPa
2. °C
3. not applicable (N/A)
M1
1. Pressure
2. Temperature
3. Quality Metric
1. kPa
2. °C
3. N/A
M2
1. Pressure
2. Temperature
3. Quality Metric
1. inHg
2. °F
3. N/A
M3
1. Pressure
2. Temperature
3. Quality Metric
1. mmHg
2. °C
3. N/A
M4
1. Pressure
2. Temperature
3. Quality Metric
1. psia
2. °F
3. N/A
BaroVUE™10 Digital Barometer
24