8. Unpack the replacement sensor card, avoiding static discharges to the element by touching
the package before touching the element.
9. Either hold the new sensor card by the top of the package (the other end to the gold pins)
or use a pair of fine nosed pliers or tweezers to grip the sensor card.
10. Carefully align and insert the sensor card in the card socket on the motherboard
11. Confirm the correct orientation and gently push the sensor card into the socket until it will
not go any further.
12. Before closing the main case, double-check that the card is inserted in the correct
13. Close the main case and secure it by pressing down on the latch and clipping it into
position.
CAUTION:
Ensure that cables and pressure tubes are clear of the case and not pinched when
closing the lid and tightening the case. This can damage the case, cables, and pressure
tubes.
CAUTION:
Do not apply extreme force when closing the lid. The main case should close with ease.
If difficulty is experienced, check for an obstruction.
BaroVUE™10 Digital Barometer
32