Advanced Graphic Data Manager
Commissioning and use during operation
55
Submenu: Measured value
correction
For determining the correction values to offset measurement section
tolerances.
Follow these instructions:
1.
Measure the current value at the lower measurement range.
2.
Measure the current value at the upper measurement range.
3.
Enter the lower and upper target and actual value.
Fig. 37:
Setup - Inputs - Universal inputs, Universal input x,
"Measured value correction" submenu
Lower correction value:
Target value:
enter the lower target value here (e.g. measuring range
0°C to 100°C: 0°C).
23000/000 to
23000/039
Actual value:
enter the lower value that is actually measured here (e.g.
measuring range 0°C to 100°C: 0.5°C measured).
23005/000 to
23005/039
Upper correction value:
Target value:
enter the upper target value here (e.g. measuring range
0°C to 100°C: 100°C).
23010/000 to
23010/039
Actual value:
enter the upper value that is actually measured here (e.g.
measuring range 0°C to 100°C: 100.5°C measured).
23015/000 to
23015/039
Correction RPT:
Rear panel temperature correction value for this universal input (only
required for thermocouples).
Note:
Only appears once the service code has been entered!
23500/000
to
23500/039
"Inputs" menu
items
Configurable parameters
(factory settings are highlighted in bold)
Direct access
code
Содержание LINAX A325
Страница 1: ...Operating instructions Video recorder LINAX A325 ...
Страница 3: ...Advanced Graphic Data Manager 3 ...
Страница 151: ...Advanced Graphic Data Manager Index 151 ...
Страница 152: ...BA00247O 09 en 03 14 71252214 FM10 ...