Number Message String
Cause
Remedy
exceeded.
caused by switching EUT on or off.
827
“Frequency error”
Frequency error during
self-test.
Correct frequency was not measured
during self-test. May be result of 801
error.
828
“Phase error”
Self test error phase
angle
Correct phase angle was not measured
during self-test. May be result of 801
error.
829
"Dc component exceed
limit"
DC value exceed limit for
the voltage mode setting.
Select AC+DC mode instead of AC
mode.
830
"Lan connection missing”
LAN Cable is not
connected
Make sure the LAN cable is connected
to a network.
831
"Duplicate IP,Set to
DHCP”
It detect a duplicate IP on
the network.
Change IP or Cycle power to connect to
DHCP or Auto IP.
832
"Duplicate IP,REM set
reject”
The IP set from remote
interface is rejected.
The IP is rejected use different IP. It is a
duplicate IP or out of reach IP.
833
“Duplicate IP,LOC set
rejec”
The IP set from front
panel keypad is rejected
The IP is rejected use different IP. It is a
duplicate IP or out of reach IP.
834
"Out of reach IP"
The IP set is out of reach
Use an IP within the sub net.
Table 16-1: Error Messages