3. Unpacking and Installation
3.1 Unpacking
Inspect the unit for any possible shipping damage immediately upon receipt. If damage is evident,
notify the carrier.
DO NOT
return an instrument to the factory without prior approval. Do not destroy
the packing container until the unit has been inspected for damage in shipment. If possible, retain the
container in the event the system ever has to be returned to the factory for either repair or upgrades.
WARNING: This power source weighs approximately
60 lbs / 28 Kg.
Obtain adequate help
when moving or installing the unit. For cabinet mounting, use rack supports to support the
weight.
3.2 AC Input Power Requirements
The i/iX Series power source has been designed to operate from a single-phase, two wire AC input
line. A protective earth connection is required as well. (PE). Available AC input voltages are from 115
to 230 V
LN
nominal.
CAUTION: Ensure the outlet used to power the 2253i/iX is rated for sufficient current.
3.3 Mechanical Installation
3.3.1 Table top
The iX/i Series AC power sources can be used free standing on a solid surface or mounted in a 19”
instrument cabinet. The units are fan cooled, drawing air in from the side and exhausting at the rear.
The back of each unit must be kept clear of obstruction and a 3” clearance must be maintained to the
rear. Special consideration of overall airflow characteristics and the resultant internal heat rise must
be considered at all times to avoid self heating and over temperature problems.