MN260 B | 45
Service Manual
7.8.7.2 Insert and secure the Battery Power and Compressor
Power Wires where shown. Use a flat blade screwdriver to
tighten and torque the screws to 4.4 in-lbs.
NOTE: Battery wire polarity indicated on the PCBA. Red is (+) and
Black is (-).
battery
power
wires
compressor
power wires
Figure 7-62. Secure Wires
7.8.7.3 Connect the Thermistor Connector to the PCBA
7.8.7.4 Secure Ferrites to the Compressor Wires.
Figure 7-63. Secure Ferrites
7.8.7.5 Connect the Hall Effect Connector to the PCBA
NOTE: Ensure there is enough slack to the Hall Effect Wire
Figure 7-64. Connect
7.8.7.6 Connect the Fan Connector to the PCB and connect flow
tubing.
Figure 7-65. Fan Connector
Содержание SAROS 3000
Страница 56: ...Service Manual 56 MN260 B NOTES...
Страница 57: ...MN260 B 57 Service Manual NOTES...