7.7.2.8 Disconnect the Breath Detect Tube.
Figure 7-14. Tube Disconnect
7.7.2.9 Remove the Extended Top Cap (Part# SP9756V2-SEQ)
and guide AC Power Cable under the frame of the PCB Assem-
bly. Remove and set aside speaker foam.
Figure 7-15. Inlet Cap Removal
disconnect
Guide
Wires
7.7.3 REMOVE PCB ASSEMBLY
NOTE: Always use ESD Controls when handling the PCB Assembly.
7.7.3.1 Cut zip tie and remove ferrite clamp from Battery Power
Wire.
Figure 7-16. Removal of Ferrite from Battery Power Wire
7.7.3.2 Loosen the Battery Power Wire screws and remove wires.
7.7.3.3 Disconnect the Compressor Thermistor wires.
7.7.3.4 Disconnect the Halls Effect wire.