MN260 B | 29
Service Manual
7.7.2.4 Move sleeve ring aside and cut green cable tie and dis-
connect the Oxygen Outlet Tube from the Flow Valve.
Figure 7-10. Oxygen Outlet Tube Disconnect.
7.7.2.5 Remove tape (Part# 2696-SEQ) from Front Panel Flex
Cable and Top Mounting Struts (Part# 9778-SEQ). Lift up lock-
ing connector and disconnect the Front Panel Flex cable.
Figure 7-11. Front Panel Flex Cable Disconnect
7.7.2.6 Disconnect the Ambient Air Thermistor Connector and
9V Battery Connector.
Figure 7-12. Thermistor Connector and 9V Battery Connector
7.7.2.7 Loosen the screws for the AC Power Connector and
remove the wires.
Figure 7-13. Inlet Cap Power
oxygen
outlet
tube
front panel
flex cable
9V Battery
connector
ambient
thermistor
connector
loosen
Содержание SAROS 3000
Страница 56: ...Service Manual 56 MN260 B NOTES...
Страница 57: ...MN260 B 57 Service Manual NOTES...