MN260 B | 53
Service Manual
9.0 SERVICE RECORDS & RETENTION
9.1 RECORD HOURS OF OPERATION AND SOFTWARE VERSION
To help maintain the SAROS, you may obtain the Total Hours of Operation and software version numbers for the Control Board.
To obtain system information, follow these steps:
Press the “Utility” Button on the control panel and advance to the following display mode for each time the icon is depressed:
• Tactical Mode
• Alarm Code
• Pulse Sensitivity
• LCD Contrast
• Tactical Volume
• Tactical Brightness
• Normal Brightness
• 9-volt Battery Status
• Hour of Operation
• System Software Version
• Motor Software Version
Total Hours of Operation
Alarm Code
Pulse Sensitivity
LCD Contrast
Tactical Volume
Tactical Brightness
Normal Brightness
9-Volt Battery Status
Hour Meter
System Software Version
Motor Software Version
Table 9-1. Log the Results
Type of protection against electric shock:
Class II Protection from electric shock is achieved by double
insulation.
Degree of protection against electric shock:
Type BF
Equipment providing a particular degree of
protection against electric shock regarding
1) allowable leakage current;
2) reliability of protective earth connection (if present).
Not intended for direct cardiac application.
Independent testing for Medical Electrical Equipment Stan-
dard:
Tested to be in compliance with
• IEC 60601-1 edition 3.1: Medical Electrical equipment -
Part 1 General Requirements for safety.
• CAN/CSA C22.2 No. 60601-1-8 edition 2.1, Medical
Electrical Equipment – Part 1: General Requirements for
Safety
Protection against potential electromagnetic or other interfer-
ence between the equipment and other devices.
• Tested to be in compliance with EN 60601-1-2/IEC
60601-1-2: 4th Edition.
• Tested to be in compliance with RTCA/DO160
Section 21 Category M.
• CISPR 11 / EN 55011 Class B Group 1, “Industrial,
Scientific, and Medical (ISM) Equipment”
• FCC Part 15, Subpart B – Class B Unintentional Radiators
IP33 – Protection against ingress of tools, thick wires, etc. >
2.5mm. Water falling as a spray at any angle up to 60° from
the vertical shall have no harmful effect.
Method of cleaning and infection control allowed: Please refer
to “Cleaning, Care, and Proper Maintenance” section of this
FreeStyle User Manual.
Degree of safety of application in the presence of flammable
anesthetic gases: Equipment not suited for such application.
Mode of operation: Continuous duty.
8.0 CLASSIFICATION