Advanced settings
Fig. 1
7
:
"Reject point" screen using a function key.
7.3.4 Sniffing reject set point
The sniffing reject set point defines the acceptance set point for parts that are "accepted/
rejected" in a sniffing test:
– Measured leak rate ≤ reject set point: part accepted
– Measured leak rate > reject set point: part rejected.
7.3.5 Probe clogged set point
The purpose of this set point is to check that the sniffer probe (accessory) is operational.
When the measured leak rate is lower than the set ’probe clogged’ set point, the operator
receives a message to check the probe
(See
Sniffer probe
Operating instructions)
.
● The set point unit is the unit set for the detector.
Block the end of the sniffer probe from time to time with your finger to check that the
leak rate is going down. If not, the probe may be clogged or drilled. Do not block the
end for too long: if it is the case, a error message displays.
For quick access from the control panel, set a function key for [Reject Point]
From the "Settings" screen, press [Set Points] [Sniffing Set Points].
Reject point
Set the set point value.
For quick access from the control panel, set a function key for [Reject Set Point]
From the "Settings" screen, press [Set Points] [Sniffing Set Points].
Probe clogged
Set the set point value.
Fig.
18
: "Reject point" screen using a function key.