Advanced settings
7.4
Test Menu
From the "Settings" screen, press
[Test]
.
Fig.
19
:Hard vacuum test 1 and sniffing test 2 menu
7.4.1 Test method
There are 2 possible test methods:
● hard vacuum test
(see 6.4.1)
,
● sniffing test
(see 6.4.2)
(Sniff
er
probe required)
.
Switching from ’Hard Vacuum test’ to ’Sniffing test’
After modifying the settings, a transition duration of < 3 min during which the test can be
performed but calibration is not possible.
Switching from ’Sniffing test’ to ’Hard Vacuum test’
After modifying the settings, a transition duration of 30 s during which neither the test nor
the calibration can be performed.
Fig. 2
0
:
"Method" screen using a function key
1
2
From the "Settings" screen, press [Test].
Method
Select the test method.
– For the hard vacuum test, set the test mode
.
NOTICE
Limit of operation
Make sure that the parts or chambers connected to the inlet of our products withstand
a negative pressure of 1·10
3
hPa in relation to atmospheric pressure.
For quick access from the control panel, set a function key for [Method]
50
0870231897_0001_IM_en