89
of 175
Slot
Index Description
Value
Meaning of value
Type
Len R/W
44
35
Set the "stop" number of
measured values
0 bis 5000
Integer value
U16
2
RW
44
36
Set bend-up factor
between
-9999999.0 and
9999999.0
Float value
Float according to
IEEE754
FLT
4
RW
7.4.16 Evaluation window 1 (Slot 45)
Slot
Index Description
Value
Meaning of value
Type
Len R/W
45
0
Not possible
-
-
X
45
1 .. 9 Reserved
-
-
X
45
10
Window 1 off/on
0
1
off
on
U16
2
RW
45
11
Window 1 limit Xmin
Note
: At the end, entry
must be adopted
through index 15.
between
-9999999.0 and
9999999.0
Float value
Float according to
IEEE754
FLT
4
RW
45
12
Window 1 limit Xmax
Note
: At the end, entry
must be adopted
through index 15.
between
-9999999.0 and
9999999.0
Float value
Float according to
IEEE754
FLT
4
RW
45
13
Window 1 limit Ymin
Note
: At the end, entry
must be adopted
through index 15.
between
-9999999.0 and
9999999.0
Float value
Float according to
IEEE754
FLT
4
RW
45
14
Window 1 limit Ymax
Note
: At the end, entry
must be adopted
through index 15.
between
-9999999.0 and
9999999.0
Float value
Float according to
IEEE754
FLT
4
RW
45
15
Window 1 copy limit
Note
: Values entered
into indices 11, 12, 13,
14 are adopted
EVENT!
Writing an arbitrary byte
initiates action
U8
1
WO
45
16
Window 1 entry left
Note
: At the end, entry
must be adopted
through index 24.
0
1
no
yes
U16
2
RW