60
of 175
Slot
Index Description
Value
Meaning of value
Type Len R/W
32
12
Fault status of the internal
serial communication
0x00000100
Invalid configuration
U32x 4
RO
32
12
Fault status of the internal
serial communication
0x00000200
Scaling fault
U32x 4
RO
32
12
Fault status of the internal
serial communication
0x00000400
No valid measurements
are available
U32x 4
RO
32
12
Fault status of the internal
serial communication
0x00000800
A/D converter overdriven
U32x 4
RO
32
12
Fault status of the internal
serial communication
0x00001000
Fault reading from
EEPROM
U32x 4
RO
32
12
Fault status of the internal
serial communication
0x00002000
Overdrive resulting from
scaling
U32x 4
RO
32
12
Fault status of the internal
serial communication
0x00004000
Reading out the
measurement curve was
interrupted by the
beginning of a new
measurement
U32x 4
RO
32
12
Fault status of the internal
serial communication
0x00008000
Invalid envelope limits
U32x 4
RO
32
12
Fault status of the internal
serial communication
0x00010000
The calibration has not
worked
U32x 4
RO
7.4.4 Minimal setup menue (Slot 33)
Slot
Index Description
Value
Meaning of value
Type Len R/W
33
0
Not possible
-
-
x
x
x
33
1 - 9 Reserved slots
-
-
x
x
x
33
10
Set program number
0 ... 31
0…127
In the standard device
In the corresponding
device version
U16
2
RW
33
11
Program name
Program name
STR
20
20
RW
33
12
Reset statistics of
measurement program
0 ... 31
0…127
In the standard device
In the corresponding
device version
EVENT! Selection through
writing the program
number
U16
2
WO
33
13
Reset statistics in all
measurement programs
EVENT!
Writing an arbitrary byte
initiates action
U8
1
WO