Slot
Index Description
Value
Meaning of value
Type
Len R/W
42
19 ...
54
Comment on slots 19
… 54:
Reading these entries
only makes sense if the
data has been read from
the sensor beforehand
(indices 10, 11, 12)
It is also possible to
write sensor data here.
A certain risk is hidden
here: If data that is not
appropriate for the
sensor has been written,
the results of the sensor
measurements can no
longer be read correctly!
In cases of doubt, the
data should not be
written here manually,
but should instead be
read out of the sensor
through indices
10/11/12.
After all the sensor data
for the channel
concerned has been
written, they still have to
be transferred into the
device through a write
access to indices
55/56/57, and then
adopted by the device's
internal FPGA through
write access to indices
16/17/18.