Troubleshooting with messages
19.4
Messages
: maintenance required
→
If the message displayed on your product is not explained in the Operating Instructions, contact Bürkert.
19.4.1
Message
Pair cell with sensor in maintenance menu
Product-status symbol
Bit status of the PDO pH-state
bit 1 = 1
Possible cause
•
The measuring cell has been replaced.
What to do?
→
Pair the new measuring cell with the product. The detailed pro-
cedure is described in the document "Service Instructions Type
MS01 MS02 MS04". The Service Instructions is available at
.
19.4.2
Message
Calibration date has expired
Product-status symbol
Bit status of the PDO pH-state
bit 1 = 1
Possible cause
•
The calibration date is due.
What to do?
→
Calibrate the product. Refer to chpt.
19.5
Messages
: informations
→
If the message displayed on your product is not explained in the Operating Instructions, contact Bürkert.
19.5.1
Message
Transferable memory is not accessible
Product-status symbol
Bit status of the PDO pH-state
-
Possible cause
•
No memory card is plugged in the product.
What to do?
→
If a memory card is needed, insert a memory card. Contact your
Bürkert sales office to buy a new memory card.
→
If no memory card is needed, valid the message by clicking on
Acknowledge.