46
pH sensor - Diagnostics
micro-USB
connector
jack male-connector
jack female-
connector
AC/DC adapter
5-pin M12 female
connector
Fig. 14:
Electrical connection parts of the USB-büS-interface set with article number 00772426
→
If the Type MZ15 is connected to one or the other end of büS, then set the termination-resistance switch
of the büS stick to ON. Else set the termination-resistance switch of the büS stick to OFF.
→
Insert the büS stick into a USB port of the PC.
→
Wait until the Windows pilot of the büS stick has been completely installed on the PC.
→
Connect the AC/DC adapter to the power supply.
→
Start the Bürkert-Communicator software.
→
Click on in the Bürkert-Communicator software to establish the communication between the Bürkert-
Communicator software and the product. A window opens.
→
Select
büS-Stick
.
→
Choose the port
Bürkert büS Stick
, click on
Finish
and wait until the product symbol appears in the list
of devices.
→
In the list of devices, click on the symbol related to the product: the menu structure for the product is
displayed.
12.2
Manually calibrate the temperature sensor
The temperature measured by the sensor can be corrected with an offset value.
→
Make sure that the login user level is
Advanced user
. Refer to chpt.
→
Go to device
pH sensor MS01
function
pH sensor
tab
Maintenance
.
→
Select
Temperature calibration
.
→
Select
Offset
and enter a temperature value in °C. Validate with
.
The new offset value is used by the product.
The offset value is updated in the
Diagnostics
tab.