41
pH sensor - Diagnostics
→
If necessary, unscrew the termination resistance from the Type 8905 system and screw the termination
resistance on the Y junction. Refer to the Operating Instructions of the Type MZ15.
Start the calibration procedure:
Waste bottle
Bottle with the buffer
solution
Fig. 11:
Handheld calibration-and-cleaning tool: position of the bottle with the buffer solution and position of the
waste bottle
→
Screw the bottle with the first buffer solution as shown in Fig. 11
. Refer to the Operating Instructions of
the Type MZ15.
→
Make sure that the login user level is at least
Advanced user
. Refer to chpt.
→
Go to device
pH sensor MS01
function
pH sensor
tab
Maintenance
.
→
Select
pH calibration
2 point
.
The product-status indicator flashes.
Do the calibration procedure:
→
Keep the Type MZ15 handheld calibration-and-cleaning tool upright during calibration procedure
to make sure that both of the conditions are met:
-
The buffer solution flows correctly through the product.
-
The check valves operate correctly.
Step 1/9:
→
To let the first buffer solution flow through the product, press the actuator button of the Type MZ15.
→
Select
Next
.
Step 2/9:
→
Select
Input value of buffer solution 1
.
→
Enter the pH value of the first buffer solution.
→
Validate with
.
→
Select
Next
.
English
Type MS01