43
pH sensor - Diagnostics
Calibration successful
If the calibration has succeeded, you have two options:
-
Accept the new offset value and the new slope value. Select
Save
and go to the step 9/9.
The new offset value and the new slope value are displayed.
The date of the last calibration is updated. See chpt.
-
Reject the new offset value and the new slope value. Select
Cancel
and do a new calibration.
Message
Error: Value out of range
The calibration has failed because the calculated offset value or the calculated slope value are out of their
error ranges.
→
To find which value is out of its error range, refer to tab
Diagnostics
Slope limits
and tab
Diag-
nostics
Offset limits
.
•
If the slope value is out of its error range, then do the following procedure:
1.
Replace the measuring cell of the product. Refer to chpt.
2. Do the calibration again.
3. If the calibration fails again, then send the product back to Bürkert.
•
If the offset value is out of its error range, then do the following procedure:
1.
Replace the reference electrode by a new one. Refer to chpt.
2. Do the calibration again.
3.
If the calibration fails again, replace the measuring cell of the product. Refer to chpt.
4. Do the calibration again.
5. If the calibration fails again, then send the product back to Bürkert.
English
Type MS01