70
Maintenance
→
Moisten the seal with water. If not, the seal can be damaged.
→
Lift the folding lever of the reference electrode and turn it by a quarter turn to the left.
4. Insert the new reference electrode fully in the product.
→
Turn the folding lever by a quarter turn to the right and lower it fully in its seating.
Fig. 26:
Insertion of the new reference electrode in the product
5. Insert and lock the product back on the backplane of the system.
6.
Let the fluid flow through the system for at least 12 hours. Observing this time makes sure that the
product and the system operate properly and that the measurements are stable.
17.6
Replace the measuring cell of the product
The detailed procedure is described in the document "Service Instructions Type MS01 MS02 MS04". The
Service Instructions is available at
17.7
Replace a defect product by a new one
You can only replace a product by a product with the same article number.
→
Remove the defect product from the backplane.
→
Remove the memory card from its slot.
→
Pay attention to the insertion direction. Insert the memory card in the new product.
→
Mount the new product to the backplane.
→
Restart the new product to transfer the data from the memory card to the new product.
17.8
Transfer the product settings to another product
→
Choose the product whose settings you want to transfer. If necessary, remove the product from the
backplane.
→
Remove the memory card from its slot.
→
Pay attention to the insertion direction. Insert the memory card in the second product with the same
article number.
→
At the next product start-up, the second product adopts the product-specific data that is stored on the
memory card.