26
8285
NAMUR
signals
The analyzer is in measuring mode, an ISM sensor is connected.
The analyzer is in calibration mode. Function check is active.
The analyzer is in maintenance mode. Function check is active.
The analyzer is in parameter setting mode. Function check is active.
The analyzer is in diagnostics mode.
Function check
. The NAMUR “function check” contact is active (factory setting: Module
BASE, Contact K2, N/O contact). Current outputs as configured:
• Currently measured value:
The currently measured value appears at the current output
• Last measured value:
The last measured value is held at the current output
• Fix 22 mA:
The output current is at 22 mA
The analyzer contains an enabled SmartMedia „memory card“. During data recording
the dot in the icon flashes. Please note:
“Close memory card“ in the maintenance menu before removing the SmartMedia card.
Limit indication: Lower / upper range limit exceeded
The analyzer contains a SmartMedia card of the type “memory card to FDA 21 CFR
Part 11“. Serves for consistent recording of all operations (SW 8285-107).
Displayed in plain text, when the analyzer is controlled via PROFIBUS PA.
Only in conjunction with a PROFIBUS module.
Designates the module slot (1, 2 or 3), allowing the clear assignment of measured value/
parameter displays in the case of identical module types.
Indicates the active parameter set (The analyzer provides two parameter sets A and B.
Up to 5 sets can be added using additional functions and SmartMedia card.)
Failure
: The NAMUR “failure” contact is active (factory setting: Module BASE, Contact
K4, N/C contact). To view error message, call up: Diagnostics menu/Message list
Maintenance
: The NAMUR “maintenance request” contact is active (factory setting:
Module BASE, Contact K2, N/O contact). To view error message, call up: Diagnostics
menu/Message list
The analyzer contains a SmartMedia “update card“. You can save the current device
software or perform a software update from the SmartMedia card. Be sure to check the
configuration after the update is completed.
The analyzer contains a SmartMedia „memory card“. The card is closed and can be
removed or enabled in the maintenance menu.
Display Explanation of display icons
Содержание 8285
Страница 1: ...Modular Analysis Transmitter 8285 Instruction manual...
Страница 3: ...3 EC Declaration of Conformity 8285 Modular Process Analysis System...
Страница 94: ......
Страница 97: ...Transmetteur d analyse modulaire 8285 Mode d emploi...
Страница 99: ...3 D claration de conformit europ enne Syst me de mesure et d analyse modulaire 8285...
Страница 184: ...88 8285 Termes techniques Glossaire...
Страница 190: ......
Страница 193: ...Modularer Analysentransmitter 8285 Bedienungsanleitung...
Страница 195: ...3 EG Konformit tserkl rung Modulares Analysenme system 8285...
Страница 286: ......