34
8285
Parameter setting
Activated from measuring mode: Press
menu
key to select
menu.
Select parameter setting using arrow keys, confirm with
enter
.
Administrator level
Access to all functions, also passcode setting. Releasing or
blocking functions for access from the Operator level.
Operator level
Access to all functions which have been released at the
Administrator level. Blocked functions are displayed in gray
and cannot be edited.
Viewing level
Only display, no editing possible!
Menu selection
Return to meas
0.003 mS/cm
25.6 °C
Select:
[enter]
Lingua
Parameter setting
Return
0.003 mS/cm
25.6 °C
Viewing level
Administrator level
(All Data) adm
(Operation Data) opl
(All Data) view
Operator level
System control
Parameter sets
• Save
• Load
Overview of parameter setting
2 parameter sets (A,B) are available in the analyzer.
The currently active parameter set is read on the display.
Parameter sets contain all settings except:
Sensor type, Options, System control settings
Up to 5 parameter sets (1, 2, 3, 4, 5) are available when a
SmartMedia card (Option) is used.
Memory card (Option)
• Register logbook
• Register recorder
• Decimal separator
• Card full
• Formatting
Menu only appears with SmartMedia Card inserted.
Make sure that it is a memory card,
not an update card.
Commercially available SmartMedia cards must be formatted
before they can be used as memory card.
Copy configuration
The complete configuration of an analyzer can be written on
a SmartMedia card. This allows transferring all device settings
- except options and passcodes - to other devices with identi-
cal equipment.
Function control matrix
• Input OK2
• Left softkey
• Right softkey
Selecting the control element for the following functions:
- Parameter set selection
- KI recorder (Start/Stop)
- Favorites menu (selected diagnostics functions)
- Unical (fully automated probe controller)
Time/date
Point of meas description
Release of options
Software update
Logbook
Concentration chart
Factory setting
Passcode entry
Selecting the display format, entry
Can be called up in the diagnostics menu.
A TAN is required to release an Option.
Software update from SmartMedia card (update card)
Selecting events to be recorded
Entering own buffer set for automatic calibration
Resetting all parameters to factory setting
Editing the passcodes
Overview of parameter setting
Содержание 8285
Страница 1: ...Modular Analysis Transmitter 8285 Instruction manual...
Страница 3: ...3 EC Declaration of Conformity 8285 Modular Process Analysis System...
Страница 94: ......
Страница 97: ...Transmetteur d analyse modulaire 8285 Mode d emploi...
Страница 99: ...3 D claration de conformit europ enne Syst me de mesure et d analyse modulaire 8285...
Страница 184: ...88 8285 Termes techniques Glossaire...
Страница 190: ......
Страница 193: ...Modularer Analysentransmitter 8285 Bedienungsanleitung...
Страница 195: ...3 EG Konformit tserkl rung Modulares Analysenme system 8285...
Страница 286: ......