Sensoface
Graphical indication of sensor condition
Sensocheck must have been activated during parameter setting
Menu Display
Activate Sensocheck
Open menu selection
Select parameter setting
Enter passcode (Administrator)
Select module (e.g. “pH/ORP”)
Confirm with
enter
Select “Sensor data”.
Confirm with
enter
.
Then select “Sensocheck Ref el”
(Fig.)
Assign function and confirm with
enter
.
Menüauswahl
Return to meas
Select:
[enter]
Lingua
7.00 pH
25.6°C
25.6°C
Return
7.00 pH
Parameter setting (Administrator)
System control
Module 8285 FRONT
Module 8285 BASE
Module 8285 El. COND
Module 8285 pH/ORP
25.6°C
Abort
7.00 pH
Sensocheck Ref el (Administrator)
Nominal
Min
Max
Message
005.0 kOhm
003.1 kOhm
100.0 kOhm
Maint. request
Off
Failure
7.00
11.00
pH
pH
24.3°C
Sensocheck - Sensor monitoring
Module
Sensocheck function
El.COND:
Information on sensor condition
pH/ORP:
Automatic monitoring of glass and
reference electrode
The “smileys” provide information on wear and required maintenance of the
sensor (“friendly” - “neutral” - “sad”).
OK
Содержание 8285
Страница 1: ...Modular Analysis Transmitter 8285 Instruction manual...
Страница 3: ...3 EC Declaration of Conformity 8285 Modular Process Analysis System...
Страница 94: ......
Страница 97: ...Transmetteur d analyse modulaire 8285 Mode d emploi...
Страница 99: ...3 D claration de conformit europ enne Syst me de mesure et d analyse modulaire 8285...
Страница 184: ...88 8285 Termes techniques Glossaire...
Страница 190: ......
Страница 193: ...Modularer Analysentransmitter 8285 Bedienungsanleitung...
Страница 195: ...3 EG Konformit tserkl rung Modulares Analysenme system 8285...
Страница 286: ......