MAINTENANCE AND SAFETY INSTRUCTIONS | PAGE 39 OF 47
7.1.2
PROCEDURE OF WORN-OUT TUBE AND PROJECTILE REPLACEMENT
If the applicator stops working correctly after a certain time, it is possible to change the worn-out system of tube
and projectile by using the exchangeable kit that is part of the BTL - 5000 SWT package.
Do not use damaged applicators; otherwise you may cause risk of injury to the operating staff or the client.
The packaging of the spare kit includes
:
•
1x exchangeable kit
•
1x key to unscrew / screw the cartridge screw cap
•
1x plastic covering of the exchangeable kit
•
1x set of spare O-rings
Replacement procedure:
1. Firstly, according to the chapter
Applicator kit replacement wizard
start
guide of the kit replacement.
2. Unscrew the screw cap of the shock transmitter by hand (
15
).
3. Remove the shock transmitter from the applicator (
14
) (including both O - rings).
O-rings