INSTRUCTIONS FOR USE | PAGE 20 OF 47
2.5.6
THERAPY PARAMETERS SCREEN – ERGONOMIC, STANDARD AND EXPERT MODE
This screen always opens before the start of therapy, by pressing the
diag
(
1)
or
prog
(
2
) buttons (see
Diagram
of the therapy setting process
). The screen either shows the most important parameters of the therapy – you
have selected the ergonomic operation mode, or is shows all information on the therapy – you have selected the
expert or standard operation mode. In addition, the expert mode can modify all the parameters.
The differences between the modes are apparent from the following pictures:
Ergonomic mode
Standard mode
Expert mode
The operation mode can be set in the menu of the device by pressing
menu
(4)
– for more details see the chapter
Operation mode
. The expert operation mode can be entered from any mode by the quick selection – press the
man
(3)
button.
2.5.7
SETTING NUMBER OF SHOCKS
The total number of shocks for the therapy can be adjusted on screen of the therapy parameters, even during the
therapy course. The number is set either by pressing the button - number of shocks, or quick selection – by
turning the
select (9)
knob.
With the button number of shocks you can select "endless therapy" - see the picture
The endless therapy includes an infinite number of shocks, that can be applied at the start of the therapy
(end of therapy is not limited by marginal number of shocks).
2.5.8 SETTING
THERAPY
INTENSITY
The intensity (power) of the shockwave therapy can be only adjusted on the screen of the therapy parameters,
even during the therapy course. The intensity is adjusted either by pressing the
intensity
button or quick selection
– by turning the
intensity
(
12
) knob.
To enter the quick mode of intensity setting, press the intensity knob and turn it at the same time – the adjusted
intensity changes much faster.
2.6
COURSE OF THERAPY
2.6.1
START - INTERRUPTION - END OF THERAPY
Start the therapy by pressing
time / stop
(
11
). The therapy can start only, if the channel screen displays the
therapy parameters and all the parameters are adjusted correctly. The device will signalize possible
discrepancies.