MAINTENANCE AND SAFETY INSTRUCTIONS | PAGE 37 OF 47
7
MAINTENANCE AND SAFETY INSTRUCTIONS
Service inspection and measuring of all parameters of the device and relevant recalibration must be executed at
intervals shorter than 30 calendar months. The inspection is carried out by authorised BTL customer service on
the basis user's order. The manufacturer does not guarantee for the technical parameters and safe operation of
the product, if the inspection is not executed on the defined schedule.
Keep the device clean, do not store or use it for a long time in extremely dusty environments and do not spill any
liquid onto the surface. Before using check the device and its accessories (especially cables) for mechanical or
other damages. Do not use the device if damaged!
Cleaning the surface of the device and its parts:
Clean the BTL - 5000 SWT and its parts only with a lightly damp soft cloth. Use water or a 2 % detergent solution.
Never use agents containing alcohol, ammonia, petrol, thinners, etc. To clean the device, do not use abrasive
materials as they could damage the surface of the device and e.g. deteriorate metal parts of the applicator. There
are no aseptic parts of the device and there is no need to sterilize them.
Cleaning accessories getting in contact with the client (e.g. shock transmitters):
Cleanse after each client using agents approved by the competent health officer. You can use e.g. Sekusept,
Bacilol and other. For the cables of the accessories you can use e.g. Incidur spray etc. Surface of the shock
transmitter with the cap can be washed with warm water, however for the complete removal of gel from the shock
transmitter it is necessary to unscrew the shock transmitter and clean it.
Fuse replacement
The fuse of the main unit is set on the rear panel in a round black box (
27
). Fuses of the shockwave compressor
are placed on the rear panel of the unit in a round black box (
37
). Check the accuracy of the inserted fuses during
their replacement. The main unit (
25
) and the shockwave compressor (
35
) both use different types of fuses!
Before replacement, first make sure that the mains switches of the main unit (
25
) and the shockwave compressor
(
35
) are in „
0"
position and disconnect the power cords from both units. Use a suitable screwdriver or a coin to
turn the segment of the fuse case to the left and remove the fuse. Insert a new fuse and finish the replacement
similar way, but turn it to the right.
It is prohibited to insert a fuse with different indication, than mentioned
on the sign above the fuse box.
This operation can be accomplished only by a person familiarized with this procedure!
Switching the device to different mains voltage
The witch of the main unit (
28
) is placed under the device in a round black box. The switch of the shockwave
compressor (
38
) is placed on the rear panel of the unit. Before switching over, firstly make sure that the power
switches (
25, 35
) are in the „
0"
position and disconnect the power cords from the device. Use a suitable
screwdriver or a coin to turn the segment of the switches so sign on the switch with the arrow pointing to,
correspond to the voltage of the local mains distribution system.