DEVICE MENU | PAGE 30 OF 47 

3.3 SPECIFIC 

SETTINGS 

3.3.1 SHOCKWAVE 

SETTING 

 

COMPRESSOR MODE 

It adjusts two different types of the mode of pressure control in the 
system: - quick start 

- silent mode 

When using 

quick start

, the pressure control in the system is activated during switching the device on, therefore 

the therapy start is instantaneous. We recommend using this setting especially during frequent operations. The 
disadvantage of this mode is a higher noise level of the device. 

In 

silent

 

mode

 the pressure control in the system is activated only immediately before the start of therapy - during 

entry to the manual screen of the main unit and after leaving, the pressure stops being kept in the system. That 
eliminates the noise of the device, if not in use. The disadvantage can be the prolonged start of the therapy – 
according to pressure in the system the start can be extended for about 10 to 15 seconds. We recommend using 
this setting for quieter operations. 

 

THERAPY CONTROL 

Here, you can set whether you want the therapy to be controlled by the button on the applicator (

18

) or using the 

foot pedal connected to the unit of the compressor (

40

). After selecting the type of therapy control, you can 

continue by setting the options of the applicator button mode, or the foot pedal mode (see the following chapter). 

 

APPLICATOR BUTON MODE / FOOT PEDAL MODE 

This enables setting of the three different types of modes for the applicator button, or the foot pedal (according to 
setting in the previous chapter). You can choose from: 

- AUTO 

mode 

- Continual 

mode 

- ON/OFF 

mode 

When setting the 

AUTO

 

mode

 the device automatically recognizes 

the holdoff time of the button on the applicator (

18

), or pressing the 

foot pedal (

40

) and the selected way of control. If you hold the 

button or the foot pedal for more than 1sec, then after its 
subsequent release the therapy will stop (as in the 

continual

 

mode

, see below). During the time of pressing the button, or the 

foot pedal, generation of shockwaves will be in progress.  

If the button of the applicator, or the foot pedal is pressed for less than 1sec, then after its release the device will 
be set to the option 

ON/OFF

 (see below). 

By selecting the 

continual mode

 you will set the generation of the adjusted shockwaves by only one press of the 

button on the shockwave applicator (

18

), or the foot pedal (

40

), to end the generation of shockwaves, another 

touch of the button on the applicator is necessary, or pressing the pedal. 

During activated option of the 

ON/OFF mode

 it is necessary to hold the button of the applicator (

18

), or the foot 

pedal (

40

) pressed throughout the time of the application of shockwaves. Releasing the button or the pedal stops 

the generation of the shockwaves. 

The setting is saved separately for the button of the applicator and for the foot pedal. 

Содержание 5000 Shockwave Series

Страница 1: ...PAGE 1 OF 47 BTL 5000 Shockwave Series USER S MANUAL 101IE30 07 2008EN...

Страница 2: ...are always welcome since we believe an ongoing relationship with our customers is critically important to our future product line Please call or email us your suggestions While we would like you to st...

Страница 3: ...c Keypad 15 2 5 Therapy Setting Principle 16 2 5 1 Diagram of the Therapy Setting Process 16 2 5 2 Welcome Screen and Selection of Channels Tabs and Accessories 17 2 5 3 Setting Therapy Parameters Via...

Страница 4: ...formation 29 3 2 14 Unlock Code 29 3 2 15 Service Functions 29 3 3 Specific Settings 30 3 3 1 Shockwave Setting 30 3 3 2 Applicator Kit Replacement Wizard 31 4 USER OPTIONS 32 4 1 Clients 32 4 2 User...

Страница 5: ...gram will make an easy and efficient start to the therapy You can adjust any treatment parameter manually by the simple use of the touch screen buttons Throughout the whole therapy the device informs...

Страница 6: ...f calcic deposits and better vascularisation and in its final effect it causes relief of pain Outside the client s body extracorporeally a pressure pulse with a high amplitude is generated and its ene...

Страница 7: ...a probe of the applicator and into the client s body via this probe Consequently during the treatment the end of the applicator must be in immediate contact with the skin and subcutaneous tissue 1 4...

Страница 8: ...ng can temporarily occur in the treated area loss of bodily sensation or itching can temporarily occur in the treated area momentary increase of pain can appear immediately after the treatment haemato...

Страница 9: ...liver coagulation of blood disorders polypus in the area of treatment pregnancy coagulation of blood disorders and the use of anticoagulants thrombosis tumour disease polyneuropathy acute inflammatio...

Страница 10: ...lect knob to select and set individual parameters 10 esc key to cancel selection or setting and return to the previous status 11 time stop knob to set therapy time and to start and stop therapy 12 int...

Страница 11: ...urposes only 31 com in out communication link connector between the main unit and the compressor 32 type plate showing device type manufacturer and safety precautions and warning signs 33 ventilation...

Страница 12: ...ts producing heat or containing water or other liquid on the top of the device Do not place the device close to equipment producing strong electromagnetic electric or magnetic field diathermy X rays e...

Страница 13: ...orrect mains voltage 230 V or 115 V Before the first connection of the device to the mains it is necessary to check if both the mains voltage switches 26 and 28 are in the position corresponding to lo...

Страница 14: ...e system will signalize it and lock itself into a secure mode If this situation occurs please call your authorised BTL distributor 2 4 OPERATING THE DEVICE 2 4 1 TOUCH SCREEN The touch screen can disp...

Страница 15: ...e window with the numeric keypad for the parameter with the light button on the picture the button number of shocks 2000 Set the required values and return to the previous screen by pressing the enter...

Страница 16: ...E 16 OF 47 2 5 THERAPY SETTING PRINCIPLE 2 5 1 DIAGRAM OF THE THERAPY SETTING PROCESS ENTRY SCREEN DIAGNOSIS SELECTION PROGRAM SELECTION MANU L SETTING THERAPY INTERRUPTION RUNNING THERAPY THERAPY INT...

Страница 17: ...g the channels To switch between the connected accessories press the W1 tab see the picture If both shockwave accessories are connected then the channel W1 displays the tab of accessories connected to...

Страница 18: ...3 presses O and 4 presses P The actual selected letter is displayed in the box on the right next to the buttons After finding the required diagnosis select it by pressing the enter 8 button 2 5 4 SETT...

Страница 19: ...ng will display by pressing the man 3 button All specifications of the therapy can be adjusted and possibly saved as the user program or the diagnosis By pressing the particular soft buttons the actua...

Страница 20: ...ted on screen of the therapy parameters even during the therapy course The number is set either by pressing the button number of shocks or quick selection by turning the select 9 knob With the button...

Страница 21: ...is adjusted to AUTO The device automatically recognizes the holdoff time of the button on the applicator or the foot pedal and the selected method of control If you hold the button or the foot pedal...

Страница 22: ...possible therapies and examples of areas for the shockwave application The device is always supplied with a hard copy of encyclopaedia Its electronic format is included in the device and is available...

Страница 23: ...ved for each shockwave therapy Frequency Number of shocks Intensity When saving the therapy enter Name of the diagnosis therapy will be displayed in list of diagnosis under the diag 1 button Program n...

Страница 24: ...re saved in this item of the menu are the same as the previous item included in the therapy saving menu The saved therapy will also appear in both lists the list of diagnosis and the list of programs...

Страница 25: ...umber of shocks it is possible to interrupt the therapy by pressing the time stop key or the button on the applicator To end the therapy press esc The generation of shocks will finish by expiry of the...

Страница 26: ...lation of accessories provided for faster work of the device After the installation only the serial number of the accessory is downloaded from the accessory memory and the other information is read fr...

Страница 27: ...tched on There is no further work with the device possible without entering this password As standard the device is supplied unlocked with the password switched off 3 2 2 SOUND SETTING This includes a...

Страница 28: ...and esc 10 keys 3 2 6 DATE AND TIME SETTING This allows the user to set the date and time in the device 3 2 7 LANGUAGE SETTING Preferred language of the texts displayed in the device As standard Engli...

Страница 29: ...e device serial number device type firmware version etc It also contains the information up to when the device will operate so called device validity If the operation of the device is time limited thi...

Страница 30: ...following chapter APPLICATOR BUTON MODE FOOT PEDAL MODE This enables setting of the three different types of modes for the applicator button or the foot pedal according to setting in the previous chap...

Страница 31: ...ctions Procedure of the applicator kit replacement on the device 1 Connect the original applicator to the compressor 2 Start the applicator kit replacement guide 3 Now disconnect the original applicat...

Страница 32: ...ng to certain special functions of the shockwave generator Clients User sequences User diagnoses programs Recent therapies 4 1 CLIENTS This item allows you to enter edit and delete information about t...

Страница 33: ...he screen It is very similar to the creation of a new diagnosis program see the chapter Therapy saving On the tab of the compressor you can only see those therapies that were created and belong to thi...

Страница 34: ...hand rest spins and can be any time adjusted into the desired position The anatomical shape of the handle allows the therapist to grip the holder of the applicator comfortably in different positions...

Страница 35: ...APPLICATION OF SHOCKWAVES PAGE 35 OF 47 Patellar tendinopathy jumper s knee Achyllodinia Trigger points Epicondylitis...

Страница 36: ...ocus shock transmitter 15mm 1x replaceable multi focus shock transmitter 9m 1x replaceable focusing shock transmitter 15mm 1x exchangeable kit includes a spare set of O rings 1x applicator holder 1x g...

Страница 37: ...e g Incidur spray etc Surface of the shock transmitter with the cap can be washed with warm water however for the complete removal of gel from the shock transmitter it is necessary to unscrew the sho...

Страница 38: ...defined in the chapter Technical parameters 7 1 1 SHOCK TRANSMITTER REPLACEMENT PROCEDURE The shock transmitter 14 can be replaced by another one as necessary three shock transmitters are the part of...

Страница 39: ...plicators otherwise you may cause risk of injury to the operating staff or the client The packaging of the spare kit includes 1x exchangeable kit 1x key to unscrew screw the cartridge screw cap 1x pla...

Страница 40: ...NS PAGE 40 OF 47 4 Using the key part of the packaging unscrew the screw cap of the cartridge follow the pictures 5 Remove the worn out kit from the cartridge of the applicator 6 Take out a new replac...

Страница 41: ...The replaced kit must move freely 8 After the insertion of the kit firmly screw the screw cap of the cartridge back using the key 9 When replacing the applicator kit we recommend replacing the O ring...

Страница 42: ...ical mains connected to the device must be performed according to the existing valid standards IEC 364 and must be certificated according to these standards If you are not sure that the power mains su...

Страница 43: ...device from a cold environment into warm wait before plugging it into the power mains until the temperature levels 2 hours minimum Before starting the therapy check whether all the adjusted parameters...

Страница 44: ...y 50 Hz to 60 Hz Equipment protection class II according to IEC 536 External exchangeable fuse 1xT1 6A 250V tube safety fuse 5 x 20 mm according to IEC 127 2 main unit 2xT6 3A 250V tube safety fuse 5...

Страница 45: ...ric device Part 1 EMC requirements and examinations IEC 60601 1 2 ISM radio interference characteristics EN 55011 EMC electrostatic discharge IEC 61000 4 2 EMC radiated electromagnetic field IEC 61000...

Страница 46: ...any means incl electronic mechanic photographic or other records without previous approval from BTL Industries Limited BTL Industries Limited operates a policy of continuous development Therefore it...

Страница 47: ...omical pictures x Predefined diagnosis 13 13 Preset programs 13 13 Language versions x x Sound schemes x x Colour schemes x x Recent therapy x x Electrotherapy upgrade x Laser therapy upgrade x Magnet...

Отзывы: