T419S Indicator Service Manual
37
6.2
Scale menu
3.
Repeat step
but increase
Avg
by 10.
Keep repeating steps
and
3
until the scale is stable or you’ve tried an
Average value up to 80. If the scale is still not stable go to step
.
4.
Setting the
Const
to higher values increases the filtering effect.
To do this: Set
T-Hold t
o 0.0,
Const
to 1, and
Avg
to 80. Check the stability of
the scale by exiting to normal weight mode and observe the Center of Zero
annunciator. If it is on all the time your scale is stable. If the Center of Zero light
blinks, more filtering is required. Go to step
5.
Repeat step
but increase the
Const
by 1. Keep repeating steps
and
the scale is stable or you’ve tried the entire range of Constant (10). If the scale
is still not stable, decrease your display update rate and start over at step
using the new, slower display rate.
6.
After the
Const
value is established you may wish to lower the
Avg
value to
improve display response time.
7.
After a final value for
T-Hold,
Const
and
Avg
has been set, enter the
T-hold
value established in step
. If this value is too small your scale will act as if the
filtering is off or not working.
Increase the
T-hold
value until your scale stabilizes.
If the Threshold value is too high, your scale will react slowly to weight
changes.
When filtering is properly adjusted the scale will be stable at zero and will
rapidly display a stable test weight value.
6.2.7 Ranges
Use this to set the following ranges.
0-Rnge
Use this to set the percentage of capacity which can be zeroed off the
scale by pressing the
ENTER
key.
O-Cap
This stands for over capacity. Use this to configure the over/under capacity
function of the scale. See the items below.
Basis
Use this to choose what the over and under capacity function
is based on.
Percent
Use this to choose to base over/under capacity on a percent.
Div
Use this to choose to base over/under capacity on number of
divisions.
O-Load
This stands for overload. Once you’ve picked the basis for an
over capacity condition, use this item to set the value that
triggers the overload condition.
U-Load
This stands for underload. Use this item to set the value that
triggers the underload condition.
Содержание T419S
Страница 1: ...AWT35 501788 Issue AC T419S Indicator Service Manual original instructions AMERICAS CONFIDENTIAL ...
Страница 2: ...BTEK_T419S_s_en_501788 book 2019 B TEK All rights reserved ...
Страница 118: ...118 T419S Indicator Service Manual ...
Страница 130: ...130 T419S Indicator Service Manual ...
Страница 134: ...134 T419S Indicator Service Manual Last 28 Temp 28 ...
Страница 135: ......
Страница 136: ...B TEK Scale LLC 1510 Metric Ave SW Canton OH 44706 3088 ...