24
T419S Indicator Service Manual
5
Diagnostics menu
3.
To test input 2, jumper pins 1 and 3 of the I/O connector on the indicator …
The second digit becomes
2
until the jumper is removed.
4.
To test input 3, jumper pins 1 and 4 of the I/O connector on the indicator ….
The third digit becomes
3
until the jumper is removed.
5.
Press
ENTER
…
Inputs
is displayed.
5.7 Outputs
The output test is used to verify if external relays or lights (etc.) connected to TB2 are
properly wired and functioning properly.
Follow the steps below to perform the output test.
1.
With
Outputs
displayed, press
SELECT
…
Output1
is displayed.
2.
Press
SELECT
…
o.1-Off
is displayed.
3.
Press
or
UNITS
to toggle the output
on
(
o.1-on
) and repeat to turn it
Off
.
Output 1 will be toggled on and off as you press the keys. This is shown
by the annunciator (
SP1)
on the display turning on and off.
4.
Press
ENTER
or
F1
to stop the test …
Output1
is displayed.
5.
Press
UNITS
to go to the next output. Repeat the steps to test output 2 and 3.
6.
When finished, press
TARE
…
Outputs
is displayed.
5.8 Options
Use this to test the various installed option cards.
Bus x
Refers to Bus 1 or Bus 2, where the option card is attached.
Card x
Refers to the position of the S1 switches on the option card.
See
CAUTION: Be sure to take proper precautions to ensure material controlled by
the scale outputs will not create a hazardous condition during an output test.
Содержание T419S
Страница 1: ...AWT35 501788 Issue AC T419S Indicator Service Manual original instructions AMERICAS CONFIDENTIAL ...
Страница 2: ...BTEK_T419S_s_en_501788 book 2019 B TEK All rights reserved ...
Страница 118: ...118 T419S Indicator Service Manual ...
Страница 130: ...130 T419S Indicator Service Manual ...
Страница 134: ...134 T419S Indicator Service Manual Last 28 Temp 28 ...
Страница 135: ......
Страница 136: ...B TEK Scale LLC 1510 Metric Ave SW Canton OH 44706 3088 ...