T419S Indicator Service Manual
25
5.9
Logs
5.9 Logs
Logs
These are logs of overload and underload conditions experienced by the
scale. You can print them or clear them from memory.
Choose to print the log from Port 1, Port 2, Port 3 or USB.
Clear
Choose to clear the log from memory.
5.10 BSQ
The BSQ menu item provides the digital frequency information for the crystals on the
QDT (Quartzell Digital Transducer).
Scale 1
through
Scale 4:
Select which Scale number the BSQ is assigned,
Scale 1
through
Scale 8
.
Comprss
This stands for compression frequency.
Tension
This stands for tension frequency.
The nominal value for the tension and compression frequency is 47,200 ± 2000 kHz.
The tension and compression frequencies should each be as stable as the other and
within 10% of each other. As weight increases the tension frequency should increase
and the compression frequency should decrease.
5.11 DigJbox
This item gives you information about the ZB210 digital junction box. For more
information on the ZB210 refer to the ZB210 Installation/Service manual, PN AWT35-
501806.
The T301 item is not available in the US market.
This completes the Diag menu. To exit the menu, see
.
The logs report will be a printout of overloads and underloads that have occurred.
Содержание T419S
Страница 1: ...AWT35 501788 Issue AC T419S Indicator Service Manual original instructions AMERICAS CONFIDENTIAL ...
Страница 2: ...BTEK_T419S_s_en_501788 book 2019 B TEK All rights reserved ...
Страница 118: ...118 T419S Indicator Service Manual ...
Страница 130: ...130 T419S Indicator Service Manual ...
Страница 134: ...134 T419S Indicator Service Manual Last 28 Temp 28 ...
Страница 135: ......
Страница 136: ...B TEK Scale LLC 1510 Metric Ave SW Canton OH 44706 3088 ...