28
T419S Indicator Service Manual
6
ADMIN Menu
6.1 Calibrate
Under
Calib
you first must choose which scale to calibrate. If there is only one scale
connected,
Scale 1
is the only choice displayed. There can be up to eight scales.
Highlight the scale you want to calibrate and press
ENTER
.
The list of procedures available appears, as shown in
.
Figure 6.4 Calibration menu
6.1.1 Zero procedure
Highlight
Zero
to set the zero point for the scale and press
ENTER
CAL.ZERO procedure
1.
Highlight
Cal.Zero
and press
ENTER
…
The
CALIBRATE
window appears.
• Scale 1-8
• Zero
• Span
• Linear
• Build.Up
• Input
• Gravity
• Display
• Cal.Unit
• DigJBox
Calib
- See
- See
- See
Linearity (Linear) procedure on page 30
- See
- See
- See
- See
- See
- See
- See
Calib
• Cal.Zero
• Temp
• Last
Zero
• Scale 1-8
Set the zero for the scale using one of the following three methods.
Cal.Zero
Use this to record the zero point.The word
CALIBRATE
on
the display denotes the fact you are in the calibration
procedure. See
below.
Temp
This is an alternate zeroing procedure. Use this when the
product weight on the scale, such as in a tank or vessel,
appears to be inaccurate but cannot be removed to establish
a no-load condition. See
below.
Last
This is an alternate zeroing procedure. Use this if certified
test weights placed on the scale display a slightly inaccurate
value. See
.
Содержание T419S
Страница 1: ...AWT35 501788 Issue AC T419S Indicator Service Manual original instructions AMERICAS CONFIDENTIAL ...
Страница 2: ...BTEK_T419S_s_en_501788 book 2019 B TEK All rights reserved ...
Страница 118: ...118 T419S Indicator Service Manual ...
Страница 130: ...130 T419S Indicator Service Manual ...
Страница 134: ...134 T419S Indicator Service Manual Last 28 Temp 28 ...
Страница 135: ......
Страница 136: ...B TEK Scale LLC 1510 Metric Ave SW Canton OH 44706 3088 ...