27
©2009,1996 BRUNO INDEPENDENT LIVING AIDS, INC.®
SRE-1550 02-09-2009
LEARNING THE REMOTE INFRARED TRANSMITTER
(not necessary when installing unit for the first time)
SW3 switch
9V battery
(access panel
removed)
LEARN/ERASE
button
ELECTRICAL
The operating channel of the two (2) infrared
transmitters included with the SRE-1550 is
pre-set at the Bruno factory.
Reasons for relearning transmitters:
•
there are multiple units in the same location;
• you have to replace transmitt
ers.
To relearn a transmitter:
1. Turn the circuit breaker OFF.
2. Remove the CARRIAGE COVER to expose the
circuit board.
3. Locate the LEARN/ERASE button on the circuit
board (see left).
4. On one of the IR TRANSMITTERS, remove the
battery access panel.
NOTE: If the transmitter is mounted to a wall,
unscrew the (2) mounting bracket screws, turn
the transmitter over and remove the battery
access panel
.
5. On the TRANSMITTER board, locate the switch
labelled "SW3" (see left).
6. Change the c
onfiguration for switches 1 and 2.
Note: There are four possibl
e configurations:
• 1 up, 2 down
(default manufacturer's setting)
• 1 up, 2 up
• 1 down, 2 up
• 1 down, 2 down
7. Once you have changed the switch positions:
• turn on the CARRIAGE circuit breaker;
• wait until you hear a BEEP.