ELECTRICAL
Changing the PCB Diagnostic Mode
1. Turn the circui
t breaker on the carriage to `OFF’.
2. Remove the left carriage cover.
3. The unit is shipped in the Multi-User Diagnostic Mode. Changes are made via
the Number 1 and Number 2 positions on the 4-Ganged DIP Switch.
4 GANGED DIP SWITCH
Chirp
0.25 seconds
Short Beep
0.50 seconds
Long Beep
1.50 seconds
Pause
1.00 second
DIP SWITCH POSITION
DIAGNOSTIC MODE #1
#2
Multi-User OFF
OFF
Single-User OFF
ON
Quiet ON
OFF
Battery Warnings Only ON
ON
AUDIO REFERENCE
NOTICE:
Before touching anything inside the carriage
assembly, ground yourself by touching an unpainted
metal surface on the unit such as an exposed bolt,
or one of the mounting screws on the
electrical panel. While you work, periodically touch
an unpainted metal surface to dissipate any static
electricity that could harm internal compo
nents.
CIRCUIT BOARD DIAGNOSTICS
Also located on the 4-ganged DIP switch are switches #3 Installation and #4
Coast Delay. The coast delay option (Switch #4) has been provided in cases of
interference which may cause intermittent operation. The normal setting is 600
mSec. of coast. Should the unit lose the remote call/send signal, this can be
increased to 900 mSec. by moving Switch #4 to the `ON’ position.