Alignment
VÅNTEC-1 Detector User Manual
8 - 2
M88-E01072
For additional information on running a mea-
surement, see Section 9.
NOTE: The detector must be set in the Details
tab of DIFFRAC
plus
COMMANDER before
selecting fixed PSD scan.
2.
After running the baseline scan, measure
the ratio of the intensity of the main peak to
the valley between the K
α
1
and K
α
2
peaks
(see Figure 8.1).
NOTE: The increment is dependent on PSD-
Calib and the measuring diameter.
Figure 8.1 - Measure the intensity ratio
Start
33.6°
Stop
36.6°
Increment
0.006°
Scan type
Fixed PSD
Near window slits
Open
Detector window slits
Open
Soller slit
2.5°
Ni filter
In
Debye slit
Out
Scan speed
30 sec. per step
Table 8.1 – Scan parameters for SRM 1976
K
α
1
K
α
2
First scan
Second scan
Third scan
Содержание VANTEC-1
Страница 1: ...V NTEC 1 Detector USER MANUAL M88 E01072 5 05 BRUKER ADVANCED X RAY SOLUTIONS...
Страница 2: ......
Страница 3: ...V NTEC 1 Detector User Manual BRUKER ADVANCED X RAY SOLUTIONS M88 E01072 5 05...
Страница 33: ...V NTEC 1 Detector User Manual System Description M88 E01072 3 19 Figure 3 17 Counterbalance positions...
Страница 34: ...System Description V NTEC 1 Detector User Manual 3 20 M88 E01072...
Страница 76: ...Hardware Installation for the D4 ENDEAVOR V NTEC 1 Detector User Manual 5 10 M88 E01072...
Страница 96: ...Calibration V NTEC 1 Detector User Manual 7 8 M88 E01072...
Страница 102: ...Alignment V NTEC 1 Detector User Manual 8 6 M88 E01072...
Страница 122: ...Appendix A Configuration Tables V NTEC 1 Detector User Manual A 4 M88 E01072...
Страница 128: ...Appendix B Spare Parts List V NTEC 1 Detector User Manual B 6 M88 E01072...
Страница 136: ...Appendix C Updating the Detector Controller Firmware V NTEC 1 Detector User Manual C 8 M88 E01072...
Страница 137: ...M88 E01072 D 1 Appendix D EC Declaration of Conformity See the figure on the next page...
Страница 138: ...Appendix D EC Declaration of Conformity V NTEC 1 Detector User Manual D 2 M88 E01072...