Hardware Installation for the D8 ADVANCE
VÅNTEC-1 Detector User Manual
4 - 18
M88-E01072
4.
Answer Yes to the confirmation message
and click Additional Settings. The follow-
ing box appears (Figure 4.15). Select the
axis that should deliver the Clock signal.
Because the logical axis TH-Detector is
sending the clock signals, this motor axis
has to be chosen. From Figure 4.13, the
“Axis” must be Driver 1. Set “Register” to
Signal Status Register and “Bit” to (bit 11)
Motor Clock - Bit 3.
Figure 4.15 - Motorized Drives section – TH-Detector –
Advanced Configuration
5.
Also check the settings of the logical axis
TH-Tube in the same way. The parameters
for “Sources for AxisOUT” must look like
Figure 4.16.
Figure 4.16 - Motorized Drives section – TH-Tube –
Advanced Configuration
6.
Save and download the configuration.
Содержание VANTEC-1
Страница 1: ...V NTEC 1 Detector USER MANUAL M88 E01072 5 05 BRUKER ADVANCED X RAY SOLUTIONS...
Страница 2: ......
Страница 3: ...V NTEC 1 Detector User Manual BRUKER ADVANCED X RAY SOLUTIONS M88 E01072 5 05...
Страница 33: ...V NTEC 1 Detector User Manual System Description M88 E01072 3 19 Figure 3 17 Counterbalance positions...
Страница 34: ...System Description V NTEC 1 Detector User Manual 3 20 M88 E01072...
Страница 76: ...Hardware Installation for the D4 ENDEAVOR V NTEC 1 Detector User Manual 5 10 M88 E01072...
Страница 96: ...Calibration V NTEC 1 Detector User Manual 7 8 M88 E01072...
Страница 102: ...Alignment V NTEC 1 Detector User Manual 8 6 M88 E01072...
Страница 122: ...Appendix A Configuration Tables V NTEC 1 Detector User Manual A 4 M88 E01072...
Страница 128: ...Appendix B Spare Parts List V NTEC 1 Detector User Manual B 6 M88 E01072...
Страница 136: ...Appendix C Updating the Detector Controller Firmware V NTEC 1 Detector User Manual C 8 M88 E01072...
Страница 137: ...M88 E01072 D 1 Appendix D EC Declaration of Conformity See the figure on the next page...
Страница 138: ...Appendix D EC Declaration of Conformity V NTEC 1 Detector User Manual D 2 M88 E01072...