10 - 5
4.
The angular accuracy peak position data
is greater than +0.01° of the expected
value.
•
Repeat the calibration of the VÅNTEC-1
detector (see Section 7).
5.
There is poor instrument resolution
(FWHM is greater than 0.065° of the 104
reflection of NIST 1976).
•
Repeat the alignment of the 2 DOF mount
for the VÅNTEC-1 detector (see Section 8).
6.
The following error appears: “Scan Out-
side Software Limits.”
•
Using Fixed Scan mode, the start-to-stop
range is determined by the configuration
and cannot exceed the geometry angle set
in Config (max 12°).
Содержание VANTEC-1
Страница 1: ...V NTEC 1 Detector USER MANUAL M88 E01072 5 05 BRUKER ADVANCED X RAY SOLUTIONS...
Страница 2: ......
Страница 3: ...V NTEC 1 Detector User Manual BRUKER ADVANCED X RAY SOLUTIONS M88 E01072 5 05...
Страница 33: ...V NTEC 1 Detector User Manual System Description M88 E01072 3 19 Figure 3 17 Counterbalance positions...
Страница 34: ...System Description V NTEC 1 Detector User Manual 3 20 M88 E01072...
Страница 76: ...Hardware Installation for the D4 ENDEAVOR V NTEC 1 Detector User Manual 5 10 M88 E01072...
Страница 96: ...Calibration V NTEC 1 Detector User Manual 7 8 M88 E01072...
Страница 102: ...Alignment V NTEC 1 Detector User Manual 8 6 M88 E01072...
Страница 122: ...Appendix A Configuration Tables V NTEC 1 Detector User Manual A 4 M88 E01072...
Страница 128: ...Appendix B Spare Parts List V NTEC 1 Detector User Manual B 6 M88 E01072...
Страница 136: ...Appendix C Updating the Detector Controller Firmware V NTEC 1 Detector User Manual C 8 M88 E01072...
Страница 137: ...M88 E01072 D 1 Appendix D EC Declaration of Conformity See the figure on the next page...
Страница 138: ...Appendix D EC Declaration of Conformity V NTEC 1 Detector User Manual D 2 M88 E01072...