Software Configuration
VÅNTEC-1 Detector User Manual
6 - 10
M88-E01072
6.5.3 Set Up Communication Parameters
(both FDC types)
Under Diffractometer, select Settings.
1.1 In the pop-up window, select TCP/IP as
the connection type.
1.2 Enter the IP address 192.168.23.2 and
the port 4251.
1.3 Click OK and close the window.
Figure 6.10 - Diffractometer > Settings pop-up window
2.
In Computer Ports, select Settings under
PSD Controller.
2.1 In the pop-up window, under TCP/IP,
enter the IP address 192.168.23.2 and
the port 4250.
2.2 Click OK and close the window.
Figure 6.11 - PSD Controller > Settings pop-up window
3.
In Computer Port Configuration, click OK to
accept the changes (see Figure 6.8).
4.
To activate the configuration, go to File >
Save and Download or click the CNF icon.
Содержание VANTEC-1
Страница 1: ...V NTEC 1 Detector USER MANUAL M88 E01072 5 05 BRUKER ADVANCED X RAY SOLUTIONS...
Страница 2: ......
Страница 3: ...V NTEC 1 Detector User Manual BRUKER ADVANCED X RAY SOLUTIONS M88 E01072 5 05...
Страница 33: ...V NTEC 1 Detector User Manual System Description M88 E01072 3 19 Figure 3 17 Counterbalance positions...
Страница 34: ...System Description V NTEC 1 Detector User Manual 3 20 M88 E01072...
Страница 76: ...Hardware Installation for the D4 ENDEAVOR V NTEC 1 Detector User Manual 5 10 M88 E01072...
Страница 96: ...Calibration V NTEC 1 Detector User Manual 7 8 M88 E01072...
Страница 102: ...Alignment V NTEC 1 Detector User Manual 8 6 M88 E01072...
Страница 122: ...Appendix A Configuration Tables V NTEC 1 Detector User Manual A 4 M88 E01072...
Страница 128: ...Appendix B Spare Parts List V NTEC 1 Detector User Manual B 6 M88 E01072...
Страница 136: ...Appendix C Updating the Detector Controller Firmware V NTEC 1 Detector User Manual C 8 M88 E01072...
Страница 137: ...M88 E01072 D 1 Appendix D EC Declaration of Conformity See the figure on the next page...
Страница 138: ...Appendix D EC Declaration of Conformity V NTEC 1 Detector User Manual D 2 M88 E01072...