78
ZTKS0156 / Z31820 / Rev.: 03
Assembling
7.
Insert the other side of the helium transfer line into the helium transportation dewar
slowly. Fast insertion of the warm transfer line causes evaporation of helium and
may cause overpressure inside the transportation dewar.
8.
Adjust a pressure of 0.1 - 0.2 bar (1.5 - 3 PSI) at the helium transportation dewar.
=> Helium transfer starts.
i
Time for cooling down to 4.2 K with helium will take 4 to 6 hours.
9.
Use the IBT temperature sensor to monitor the cooling down (see chapter Appendix
Technical Data).
10. Stop helium transfer at a helium level of 100 % at the helium level sensor and 4.2 K
at the IBT temperature sensor.
11. Remove the pumping unit and the vacuum valve (refer to
).
12. Close the stop valve at the cold head turret and remove the helium gas transporta-
tion dewar.
13. Mount the supplied connector (1) to the helium flow system and connect the con-
nector with the stop valve at the cold head flange (3).
14. Open the stop valve at the cold head flange (3).
i
If the magnet system will be energized during the next 24 h, the current lead can remain
in the current lead turret. Otherwise continue.
15. Remove the plug out of the turret cap of the helium fill-in turret.
16. Insert the auxiliary shorting plug into the helium fill-in turret.
17. Remove the current lead out of the current lead turret.
18. Insert the shorting plug into the current lead turret with the shorting plug tool respec-
ting the correct position. The shorting plug can only be inserted in one position.
19. Close the current lead turret with the baffle tube, O-ring and screw cap.
20. Remove the auxiliary shorting plug out of the helium fill-in turret.
21. Close the helium fill-in turret immediately with the plug of the turret cap.
22. Record the filling session (see Appendix, Technical Data).
23. Observe the temperature of the outer radiation shield at the External Temperature
Control Unit. The radiation shield will reach its usual operation temperature after
four days operating. Turn on the alarm from the console and the External
Temperature Control Unit if the magnet system is at set operation temperature.
Содержание RS
Страница 1: ...NMR Spectroscopy Service Manual RS ...
Страница 4: ...4 ZTKS0156 Z31820 Rev 03 ...
Страница 8: ...8 ZTKS0156 Z31820 Rev 03 Table of Contents ...
Страница 10: ...10 ZTKS0156 Z31820 Rev 03 Contact ...
Страница 14: ...14 ZTKS0156 Z31820 Rev 03 Introduction ...
Страница 36: ...36 ZTKS0156 Z31820 Rev 03 Transportation ...
Страница 86: ...86 ZTKS0156 Z31820 Rev 03 Operation ...
Страница 112: ...112 ZTKS0156 Z31820 Rev 03 Troubleshooting ...
Страница 132: ...132 ZTKS0156 Z31820 Rev 03 Disassembling ...
Страница 133: ...133 ZTKS0156 Z31820 Rev 03 A Appendix Warning Signs Figures Tables Glossary Abbreviations Technical Data Index ...
Страница 134: ...134 ZTKS0156 Z31820 Rev 03 Appendix ...
Страница 136: ...136 ZTKS0156 Z31820 Rev 03 Appendix ...
Страница 142: ...142 ZTKS0156 Z31820 Rev 03 Appendix ...
Страница 144: ...144 ZTKS0156 Z31820 Rev 03 Appendix ...
Страница 150: ...150 ZTKS0156 Z31820 Rev 03 Technical Data RS Service This side is intentionally left blank ...
Страница 156: ...156 ZTKS0156 Z31820 Rev 03 ...