Troubleshooting
6.3.4
Procedure in case of Cryogenic Refrigerator failure
In case of a failure of the Cryogenic Refrigerator the temperature of the shield T
RS
will
slowly rise to 200 K and the helium boil off to 250 ml/h. This is usual behavior.
In case of failure shorter than 3 days just restart the compressor of the Cryogenic Refri-
gerator. See remarks at chapter
”During Operation of the Cryogenic Refrigerator”
for further information.
After restarting the Cryogenic Refrigerator, at least two days are necessary to recover
the previous state of the magnet system. If the temperature and helium boil off does not
decrease in the next two days call Bruker Service.
Figure 6.5:
Temperature profile during Cryogenic Refrigerator failure
Содержание RS
Страница 1: ...NMR Spectroscopy Service Manual RS ...
Страница 4: ...4 ZTKS0156 Z31820 Rev 03 ...
Страница 8: ...8 ZTKS0156 Z31820 Rev 03 Table of Contents ...
Страница 10: ...10 ZTKS0156 Z31820 Rev 03 Contact ...
Страница 14: ...14 ZTKS0156 Z31820 Rev 03 Introduction ...
Страница 36: ...36 ZTKS0156 Z31820 Rev 03 Transportation ...
Страница 86: ...86 ZTKS0156 Z31820 Rev 03 Operation ...
Страница 112: ...112 ZTKS0156 Z31820 Rev 03 Troubleshooting ...
Страница 132: ...132 ZTKS0156 Z31820 Rev 03 Disassembling ...
Страница 133: ...133 ZTKS0156 Z31820 Rev 03 A Appendix Warning Signs Figures Tables Glossary Abbreviations Technical Data Index ...
Страница 134: ...134 ZTKS0156 Z31820 Rev 03 Appendix ...
Страница 136: ...136 ZTKS0156 Z31820 Rev 03 Appendix ...
Страница 142: ...142 ZTKS0156 Z31820 Rev 03 Appendix ...
Страница 144: ...144 ZTKS0156 Z31820 Rev 03 Appendix ...
Страница 150: ...150 ZTKS0156 Z31820 Rev 03 Technical Data RS Service This side is intentionally left blank ...
Страница 156: ...156 ZTKS0156 Z31820 Rev 03 ...