70
ZTKS0156 / Z31820 / Rev.: 03
Assembling
4.3.3.6
Mounting the Helium Flow System
Figure 4.73: Mounting the Helium Flow System - step 1
1.
Clean the O-ring of the current lead turret (1)
with ethanol.
2.
Grease the O-ring (1) with vacuum grease.
3.
Clean the sealing surface for the O-ring.
4.
Mount the O-ring (1) into the sealing sur-
face.
Figure 4.74: Mounting the Helium Flow System - step 2
5.
Mount the helium flow system (1) as shown
with the one-way valve (2) showing
backwards and the flexible side at the
helium fill-in turret (7).
6.
Check the correct position of the globes in
the quench valves (6, 9) of the helium fill-in
turret (7) and the current lead turret (8).
7.
Connect the stop valve (4) with the supplied
tube (3) with the connector at the one-way
valve (2).
Содержание RS
Страница 1: ...NMR Spectroscopy Service Manual RS ...
Страница 4: ...4 ZTKS0156 Z31820 Rev 03 ...
Страница 8: ...8 ZTKS0156 Z31820 Rev 03 Table of Contents ...
Страница 10: ...10 ZTKS0156 Z31820 Rev 03 Contact ...
Страница 14: ...14 ZTKS0156 Z31820 Rev 03 Introduction ...
Страница 36: ...36 ZTKS0156 Z31820 Rev 03 Transportation ...
Страница 86: ...86 ZTKS0156 Z31820 Rev 03 Operation ...
Страница 112: ...112 ZTKS0156 Z31820 Rev 03 Troubleshooting ...
Страница 132: ...132 ZTKS0156 Z31820 Rev 03 Disassembling ...
Страница 133: ...133 ZTKS0156 Z31820 Rev 03 A Appendix Warning Signs Figures Tables Glossary Abbreviations Technical Data Index ...
Страница 134: ...134 ZTKS0156 Z31820 Rev 03 Appendix ...
Страница 136: ...136 ZTKS0156 Z31820 Rev 03 Appendix ...
Страница 142: ...142 ZTKS0156 Z31820 Rev 03 Appendix ...
Страница 144: ...144 ZTKS0156 Z31820 Rev 03 Appendix ...
Страница 150: ...150 ZTKS0156 Z31820 Rev 03 Technical Data RS Service This side is intentionally left blank ...
Страница 156: ...156 ZTKS0156 Z31820 Rev 03 ...