Assembling
i
The thread surfaces of the house couplings have to be greased to reduce friction. But
grease on the sealing surfaces of the house couplings may enter the helium cycle and
thus degrade the performance of the Cryogenic Refrigerator.
Thus the sealing surfaces have to be protected during greasing. The easiest way to
obey this advice is to connect (one turn) the male and female part as shown in the next
figure or to use the supplied plugs of the Aeroquip® couplings. Alternatively a tissue may
be used.
Figure 4.55: Mounting the Rotary Valve and Flex Lines - step 4
10. Clean the thread (1) of the female part of the
hose coupling of the flex lines with ethanol.
11. Clean the sealing surface (2) of the female
part of the hose coupling with ethanol.
Figure 4.56: Mounting the Rotary Valve and Flex Lines - step 5
12. Grease the outside thread of the male part
with PTFE-spray.
Содержание RS
Страница 1: ...NMR Spectroscopy Service Manual RS ...
Страница 4: ...4 ZTKS0156 Z31820 Rev 03 ...
Страница 8: ...8 ZTKS0156 Z31820 Rev 03 Table of Contents ...
Страница 10: ...10 ZTKS0156 Z31820 Rev 03 Contact ...
Страница 14: ...14 ZTKS0156 Z31820 Rev 03 Introduction ...
Страница 36: ...36 ZTKS0156 Z31820 Rev 03 Transportation ...
Страница 86: ...86 ZTKS0156 Z31820 Rev 03 Operation ...
Страница 112: ...112 ZTKS0156 Z31820 Rev 03 Troubleshooting ...
Страница 132: ...132 ZTKS0156 Z31820 Rev 03 Disassembling ...
Страница 133: ...133 ZTKS0156 Z31820 Rev 03 A Appendix Warning Signs Figures Tables Glossary Abbreviations Technical Data Index ...
Страница 134: ...134 ZTKS0156 Z31820 Rev 03 Appendix ...
Страница 136: ...136 ZTKS0156 Z31820 Rev 03 Appendix ...
Страница 142: ...142 ZTKS0156 Z31820 Rev 03 Appendix ...
Страница 144: ...144 ZTKS0156 Z31820 Rev 03 Appendix ...
Страница 150: ...150 ZTKS0156 Z31820 Rev 03 Technical Data RS Service This side is intentionally left blank ...
Страница 156: ...156 ZTKS0156 Z31820 Rev 03 ...