100
ZTKS0156 / Z31820 / Rev.: 03
Troubleshooting
Continued from page before
Indicator
Possible reason
Solution
By
WARNING:
Low Temperature
Not correct helium
level warning out of
supplied NMR soft-
ware.
Helium level sensor
defective.
Replace helium level sensor
(see chapter
of the Helium Level Sensor”
on page 6-109
)
Bruker Service
Helium level at con-
stant level, no
change during days.
Helium level sensor
defective.
Replace helium level sensor
(see chapter
of the Helium Level Sensor”
on page 6-109
)
Bruker Service