44
ZTKS0156 / Z31820 / Rev.: 03
Assembling
4.3.1.3
Checking the Feedthrough Plate
4.3.1.4
Assembling the Nitrogen Tube
Bottom of Cryostat
Top of the Cryostat
Figure 4.11:
Checking and fitting the Feedthrough Plate
1.
Check through the top opening of the
cryostat that all connectors at the inner side
of the feedthrough plate are plugged and the
cables are not stressed or bent.
2.
Check the feedthrough plate is correct
assembled.
Figure 4.12: Assembling the Nitrogen Tube - step 1
1.
Mount the lower RS reduction flange (1) to
the radiation shield clips.
Figure 4.13: Assembling the Nitrogen Tube - step 2
2.
Mount the upper RS reduction flange (1) to
the radiation shield clips.
Содержание RS
Страница 1: ...NMR Spectroscopy Service Manual RS ...
Страница 4: ...4 ZTKS0156 Z31820 Rev 03 ...
Страница 8: ...8 ZTKS0156 Z31820 Rev 03 Table of Contents ...
Страница 10: ...10 ZTKS0156 Z31820 Rev 03 Contact ...
Страница 14: ...14 ZTKS0156 Z31820 Rev 03 Introduction ...
Страница 36: ...36 ZTKS0156 Z31820 Rev 03 Transportation ...
Страница 86: ...86 ZTKS0156 Z31820 Rev 03 Operation ...
Страница 112: ...112 ZTKS0156 Z31820 Rev 03 Troubleshooting ...
Страница 132: ...132 ZTKS0156 Z31820 Rev 03 Disassembling ...
Страница 133: ...133 ZTKS0156 Z31820 Rev 03 A Appendix Warning Signs Figures Tables Glossary Abbreviations Technical Data Index ...
Страница 134: ...134 ZTKS0156 Z31820 Rev 03 Appendix ...
Страница 136: ...136 ZTKS0156 Z31820 Rev 03 Appendix ...
Страница 142: ...142 ZTKS0156 Z31820 Rev 03 Appendix ...
Страница 144: ...144 ZTKS0156 Z31820 Rev 03 Appendix ...
Страница 150: ...150 ZTKS0156 Z31820 Rev 03 Technical Data RS Service This side is intentionally left blank ...
Страница 156: ...156 ZTKS0156 Z31820 Rev 03 ...