Maintenance
7.4.3
Replacement of the Rotary Valve and the Cold Head
i
For removing the cold head an applicable step is necessary. This step must be stable,
not too high and without pitfalls.
i
Do not disconnect the cable from the S2 port at the feedthrough plate during the entire
procedure.
Do not switch off the External Temperature Control Unit.
Figure 7.5:
Replacing the Rotary Valve - step 1
1.
Stop the Cryogenic Refrigerator.
The temperature will increase and the
External Temperature Control Unit will give
an audible and a visual alarm.
2.
Stop the audible alarm of the External Tem-
perature Control Unit (refer to the supplied
manual of the External Temperature Control
Unit).
3.
Remove the noise protection cover from the
cold head.
4.
Release the connection line (2) from the
cold head (1).
Figure 7.6:
Replacing the Rotary Valve - step 2
5.
Remove the noise protection cap on the top
of the rotary valve column.
Содержание RS
Страница 1: ...NMR Spectroscopy Service Manual RS ...
Страница 4: ...4 ZTKS0156 Z31820 Rev 03 ...
Страница 8: ...8 ZTKS0156 Z31820 Rev 03 Table of Contents ...
Страница 10: ...10 ZTKS0156 Z31820 Rev 03 Contact ...
Страница 14: ...14 ZTKS0156 Z31820 Rev 03 Introduction ...
Страница 36: ...36 ZTKS0156 Z31820 Rev 03 Transportation ...
Страница 86: ...86 ZTKS0156 Z31820 Rev 03 Operation ...
Страница 112: ...112 ZTKS0156 Z31820 Rev 03 Troubleshooting ...
Страница 132: ...132 ZTKS0156 Z31820 Rev 03 Disassembling ...
Страница 133: ...133 ZTKS0156 Z31820 Rev 03 A Appendix Warning Signs Figures Tables Glossary Abbreviations Technical Data Index ...
Страница 134: ...134 ZTKS0156 Z31820 Rev 03 Appendix ...
Страница 136: ...136 ZTKS0156 Z31820 Rev 03 Appendix ...
Страница 142: ...142 ZTKS0156 Z31820 Rev 03 Appendix ...
Страница 144: ...144 ZTKS0156 Z31820 Rev 03 Appendix ...
Страница 150: ...150 ZTKS0156 Z31820 Rev 03 Technical Data RS Service This side is intentionally left blank ...
Страница 156: ...156 ZTKS0156 Z31820 Rev 03 ...