Planning details
▪
Fresh and ventilation in the ceiling area, exhaust in the floor area
▪
The ideal position is determined by the position of the air conditioning system connections.
Air ducts should not blow directly on to the magnet.
▪
The filter plate for the fresh air supply is assembled in an installation frame or directly
screwed to the cage.
Measurements of filter plate
Measurements of Faraday cage ventilation filter plate
Figure 6.7: Measurements of Faraday cage ventilation filter plate
6.3
Exhaust system
The magnet may only be installed or operated with a
gas extraction system for cryogenic
gases (quench pipe)
. In the case of a quench, large volumes of cryogenic gases are safely re-
moved outside using the gas extraction system.
If a magnet with a nitrogen tank is used, the installation of an additional room ventilation open-
ing in a bottom position is recommended.
WARNING
Warning of injury by cryogenic gases
The gas extraction's outlet to the outdoors must be designed in such a way that no one is able
to come into contact with cold gases.
Содержание BioSpec 117 AVANCE III
Страница 1: ...Innovation with Integrity MRI T Site planning BioSpec 117 16 AVANCE III Version 1 7...
Страница 6: ...Contents 6 ID 182463115...
Страница 33: ...ID 182463115 33 Planning details Stray field Figure 6 3 Stray field...
Страница 72: ...72 ID 182463115 Planning details...
Страница 82: ...82 ID 182463115 Delivery and transport...
Страница 84: ...84 ID 182463115 Moving or dismantling the MR system...
Страница 92: ...Figures 92 ID 182463115...
Страница 94: ...Tables 94 ID 182463115...
Страница 101: ...ID 182463115 107 Lastpage...
Страница 102: ...Bruker Corporation info bruker com www bruker com...