5. Preventative Maintenance
CryoPod Carrier
Removing the Replacing the LN
₂
Absorbent Pads
Part Number: 257437 Rev. A
Step
Action
1.
Turn the unit ON to confirm that TDISPLAY is at room temperature.
2.
Turn the unit back OFF.
3.
Remove the lid and use the included L-wrench to remove the two screws and washers that
secure the internal basket to the liner.
NOTE:
Take care to save the screws and washers for the re-installation process.
4.
Lift the internal basket out of the unit. Thoroughly clean with mild soap and water, rinse, dry
and wipe with 70% alcohol, set aside.
5.
Carefully lift out absorbent pad assembly and discard used pads per applicable
organizational Standard Operating Procedures, local, regional and national regulations.
6.
Thoroughly clean metal conductor with mild soap and water, rinse, dry and wipe with 70%
alcohol, let dry.
7.
Open the new package of absorbent pads and retrieve two pads.
8.
Put together absorbent pads and metal conductor in the same configuration as the used
pads and plate. The metal conductor flanges should slide into the cuts in the absorbent
pads.
33