Brooks Automation
3. Installation
Part Number: 257437 Rev. A
SET Mode
Step
Action
9.
Long press
to Save once all settable elements are set. A single beep will
sound to confirm that the settings have been saved.
OR
Long press
to exit without saving new settings.
SET Mode
The SET mode allows the operator to change settings and properties of the CryoPod Carrier.
NOTE:
If no button is actively pressed for 10 seconds when in SET Mode, the system
will exit the SET Mode without saving any changes.
Default SET Mode settings
SET Mode
Default Setting
Alarm 1 (TWARNING) set temperature
-155°C
Alarm 1 ON/OFF
ON
Alarm 1 sound ON/OFF
ON
Alarm 2 (TCRITICAL) set temperature
-140°C
Alarm 2 ON/OFF
ON
Alarm 2 sound ON/OFF
ON
Date (day), Date (Month), Date (Year)
No Default Setting
Time (hr), Time (min), Time (AM/PM)
No Default Setting
ID number
Different for each system
Standby (TSTANDBY) set temperature
No Default Setting
Manually Setting the CryoPod Carrier Displays
If the settings are not manually set, the default settings will be used and "NOT SET" will be displayed
in the SET position on the LCD screen. To manually set the display follow steps
6-9
of the CryoPod
Carrier Assembly Procedure.
See "CryoPod Carrier Assembly" on page 21
.
Copyright © 2016, Brooks Automation, Inc.