6. Troubleshooting
CryoPod Carrier
Functionality Check
Part Number: 257437 Rev. A
6. Troubleshooting
Functionality Check
Ensure that every time the unit is turned ON, a beep sounds and the red back light flashes to indicate
that the unit is functioning properly. When the beep sounds and the red back light flashes, the system
has turned on without any detectable issues.
Service indicator
If the SERVICE indicator turns on, the operator must contact Brooks Automation technical support.
The SERVICE indicator will appear on the LCD screen if the unit detects an error or issue. These
include thermocouple detachment (OP), corrupted temperature correlation (TMP), stuck key (KEY),
EPROM reading error (EP), and thermocouple chip error (TIC ).
NOTE:
The SERVICE error will automatically turn off once the error or issue is
resolved.
Service and Error Codes
Error Code
Description
Solution
ALARM
ALARM text and an alarm number with a red back-light will
display if the alarm is triggered. An audio alert will also sound
for 20 minutes before automatically shutting off.
Resolve the Alarms as
needed.
See "System Alarms"
on page 30
Empty battery
indicator
A battery symbol consisting of an outline of a battery and three
internal sections indicating the charge remaining:
l
All three sections filled: >75%
l
Two sections filled: >50%
l
One sections filled: >25%
l
Battery outline <13%
Replace batteries as needed.
35